Nano-1000 Micro-Spectrophotometer
Model:
Nano-1000
Keywords:
Nano-1000 Micro-Spectrophotometer
Category:
Download:
Whatsapp:
Product Description
Is a full-wavelength micro-spectrophotometer that can provide ultraviolet/visible light absorption detection.
Suitable measure a large number of nucleic acid or protein samples.
Features:
1.10-inch touch screen, Android system, no need to connect to an additional computer.
2.Results can be obtained within 8 seconds.
3.It can detect wavelengths in the range of 190 1100nm, and the maximum detection concentration can reach 15000ng/μL (DNA).
4.Samples can be loaded for detection with either hand.
5.Built-in camera for quickly judging the formation of the liquid column.
6.Log in to the instrument software through theface recognition function.
Main technical parameters
| Model | Nano-1000 | |
| Micro Detection | Sample Volume | 0.5-2μւ |
| Light Source | Xenon flash lamp | |
| Optical Path | 0.075mm、0.2mm、1mm | |
| Detector | 2048-element linear CMOS array | |
| Wavelength Range | 190-1100nm | |
| Wavelength Accuracy | ±1nm | |
| Wavelength Resolution | ≤1.5nm (FWHM@Hg253.652nm) | |
| Absorbance Repeatability | 0.002A (1mm optical path, SD) | |
| Absorbance Accuracy | 3% (0.97 A, 302 nm, 23±2 °℃) | |
| Absorbance Range | 0.04-300A | |
| Nucleic Acid Detection Range | 2-15000ng/μL (dsDNA) | |
| Test Time | About 8s/sample | |
| OD600 Detection | Absorbance Range | 0 to 4.000 Abs |
| Absorbance Stability | [0,3) ≤0.5%; [3,4) ≤2% | |
| Absorbance Repeatability | [0,3) ≤0.5%; [3,4) ≤2% | |
| Absorbance Accuracy | [0,2)≤0.005A;[2,3)≤1%;[3,4)≤2% | |
| Fluorescence Detection | Linearity | R²≥0.995 |
| Linear Range | 5 orders of magnitude | |
| Sensitivity (Limit of Detection) | <0.1 pg/μL (dsDNA HS) | |
| Others | Dimensions | 262×367×340 mm |
| Weight | 6.6kg | |